Illustrated title card for UDI glass marking

UDI Marking on Glass: What Device Makers Need to Know

A UDI (Unique Device Identifier) is a code that identifies a medical device model and, when required, its production batch. The legal duty to assign and carry it sits with the manufacturer of the finished device, not with the supplier of a glass component. GOV.UK states that raw materials, component parts and intermediate products are not normally medical devices, because the medical purpose belongs to the finished product. What the device maker needs from a glass supplier is therefore a part that can carry or support the mark, with batch records that tie every piece back to its production lot. That is the question to settle at drawing stage, and this article shows how.

Precision Glasses
Plan Glass for Reliable UDI Marking
Precision Glasses designs and fabricates custom glass components for medical devices, helping manufacturers address marking needs at the drawing stage.

Visit

TL;DR

  • The UDI obligation falls on the finished medical device manufacturer. A glass window, cover or lens is normally a component, and it needs traceability and a mark-ready surface rather than its own UDI.
  • Direct marking on the device is required where the rules call for a permanent carrier on reusable devices that are cleaned or sterilised between uses. Exceptions exist where marking would affect safety or performance, or is not technologically feasible.
  • For glass, laser etching of a Data Matrix code is the method the guidance points towards. Dot peen is unsuitable because the stylus method needs a hard, deformable material.
  • In Great Britain a UDI is not yet a legal duty under the Medical Devices Regulations 2002, but a draft amendment would make it one. Northern Ireland follows the EU regime.
  • Specify the marking zone, code size, method and reprocessing life on the glass drawing before the first RFQ.

Table of Contents

Who owns the UDI when glass is a component

The UDI has two parts. The UDI-DI (device identifier) is fixed to a device model. The UDI-PI (production identifier) changes with lot, serial number or manufacturing date. A Basic UDI-DI sits above both and groups device models for registration. The UDI carrier is the physical form, usually a barcode or a 2D Data Matrix, together with its human-readable text.

The finished device manufacturer assigns all of these. That is why a glass cover for a diagnostic scanner, a sensor window for a patient monitor or an optical window for an endoscope does not normally get its own UDI. The component feeds the device’s technical file, and the device carries the identifier.

That does not make the glass supplier irrelevant. Three things pass from the glass supplier to the device maker and decide whether the UDI works in practice:

  1. Lot traceability. If a device UDI-PI records a lot, the device maker must be able to trace the glass in that device to its own production lot, and back to the material and process records.
  2. A surface that accepts the mark. Coatings, ceramic print, edge finishing and chemical strengthening all change how a code can be applied and read.
  3. A durable finish through reprocessing. The glass, its coatings and any mark on it have to survive the cleaning regime the device sees.

We build our service around those points: ISO 9001 certification, traceable production and batch-level traceability for the glass we make. Our glass production process page sets out how a part moves from cutting through finishing to delivery, which is the chain a device maker’s auditor will want to follow.

What applies in Great Britain and Northern Ireland

The position differs by region, and the rules are moving, so check the dates before you rely on any of this.

Great Britain. The Medical Devices Regulations 2002 currently govern devices. According to the UK government’s consultation outcome, they do not create a legal obligation to obtain or provide a UDI, and the MHRA requests UDI information at registration on a voluntary basis. The same government document sets out an intention to require UDIs before devices are placed on the market, and to require reusable devices to carry a permanent UDI carrier readable after each process.

In 2026 the MHRA published a draft amendment. A law firm’s summary of it, dated 19 May 2026, says it would require manufacturers to assign a Basic UDI-DI, UDI-DI and UDI-PI before placing a device on the GB market, to place the UDI on the device or its label and on higher packaging levels, to submit the identifiers to the MHRA registration system and to reference the UDI in technical documentation. It is a draft, so treat the detail as subject to change and watch the MHRA and GOV.UK for the final text and transition periods.

Product marking. GOV.UK’s sector guidance, published on 31 March 2026, states that medical devices in Great Britain need UKCA or CE marking, and that Northern Ireland needs CE or UKNI and CE. It also records a consultation on indefinite recognition of CE-marked devices in Great Britain that opened on 16 February 2026.

Northern Ireland and the EU. Devices placed on the Northern Ireland or EU market follow EU MDR 2017/745, whose UDI rules are in Annex VI. Most UK device makers that export to the EU already design to that annex, so it is the practical reference for the glass specification, whichever market the device first reaches.

For the wider picture on the UK framework, see our guide to UK MDR 2002 Part 4A and the 2025 amendment.

What applies in Great Britain and Northern Ireland

When the mark must go on the glass itself

The rule that drives direct marking is in EU MDR Annex VI Part C. Section 4.10 requires reusable devices to bear a UDI carrier on the device itself. For devices that need cleaning, disinfection, sterilisation or refurbishing between uses, the carrier must be permanent and readable after each process, throughout the intended lifetime. An exception applies where direct marking would interfere with safety or performance, or where it is not technologically feasible.

Two further points from the same annex shape the decision:

  • The carrier must be readable during normal use and throughout the intended lifetime (section 4.11).
  • Where the carrier on the packaging can be read or scanned through the packaging, a separate carrier on the packaging is not required (section 4.12).

Single-use devices are a different case. The EU UDI helpdesk states that single-use devices are not covered by the direct-marking requirement, so a disposable device with a glass window carries its UDI on the label or packaging.

What this means for a glass part

The question for each glass component is whether the glass is the part of the device where the mark will sit. A reusable surgical instrument with a glass window may carry its mark on the metal body. A reusable optical head or a glass-bodied item may have no better surface than the glass. Decide this with the device maker’s regulatory team early, because it changes the drawing: marking zone, flatness there, coating keep-out areas and edge condition near the code.

A worked illustration: a reusable handheld scanner with a glass cover is reprocessed with a disinfectant wipe after every patient. The device manufacturer decides the housing is the right place for the carrier, so the glass needs only a lot reference in the technical file. A second manufacturer’s reusable optical probe has no housing area large enough, so the carrier must go on the glass. That glass drawing now needs a defined clear zone for a Data Matrix, a method, and a reprocessing life against which the mark has been validated. The same glass material, two different specifications.

Reusable optical probe marked on glass

Marking methods for glass

The IMDRF UDI Application Guide splits direct marking into intrusive methods (dot pin, etching, direct laser marking) and non-intrusive methods (cast, forge or mould, laser bonding, stencil, permanent adhesive label). It refers manufacturers to ISO/IEC TR 24720 for choosing a method by material, and names ISO 28219 and ISO/IEC TR 29158 as other useful standards.

Glass narrows that list quickly:

MethodSuitability for glassWhy
Dot peenUnsuitableGS1 Healthcare Japan’s guide states the stylus method requires hardness and deformability, so it is unsuitable for glass.
Laser etchingRecommended directionThe same guide recommends laser etching for small instruments to maintain reading performance.
Printed or ceramic-printed codePossible on a stable, fired surfaceA fired ceramic print is a durable, high-contrast option where the code sits on a non-contact zone. Reprocessing resistance still has to be validated.
Adhesive labelNot a permanent carrierIt suits packaging, not a mark that must survive every reprocessing cycle.

The main risk with any intrusive method on glass is damage to the part itself. The IMDRF guide asks manufacturers to analyse the effect of the substrate and chosen technology, and to ensure direct marking does not negatively affect device stability, biocompatibility or effectiveness. On glass this is a strength question as much as a legibility one: a mark placed in a high-stress zone of a chemically strengthened or toughened part can matter. For that reason the order of operations belongs on the drawing. Our guide to choosing the right approach for laser marking on glass covers the trade-offs, and our pages on chemically strengthened glass and toughened glass explain why marking location and strengthening stage need to be agreed together.

Code choice and size

ISO 28219:2009 specifies that only matrix symbologies such as Data Matrix should be used for direct part marking. It sets a minimum overall symbol grade of 1.5 ©, measured with a 0.20 mm aperture at 45° incidence. It also asks for human-readable interpretation in uppercase characters at least 1.25 mm high, with 2 mm recommended.

Size matters because the X-dimension, the width of a single module in the code, drives the reader you need. GS1’s direct marking guide warns that smaller X-dimensions affect reading performance and require special scanners or imagers. It also advises against mixing different marking methods in one reading environment. For a small optical window, the code size you can fit sets the scanner you must buy, so settle both together.

Keeping the mark readable for the device’s life

The mark has to outlast the device’s cleaning regime. The IMDRF guide asks manufacturers to ensure the UDI stays readable for the expected service life and to consider transport, storage and handling.

Its Appendix F gives three practical findings:

  • Highly alkaline cleaners, common in Europe, make direct marking fade faster over the life cycle than the neutral cleaners common in North America.
  • Too little laser engraving depth leads to faster fading.
  • A properly validated laser mark with a Data Matrix remained readable after at least 500 reprocessing cycles.

For a device sold into UK and EU hospitals, that makes the cleaning chemistry part of the specification. Write the detergent, the temperature, the number of cycles and the sterilisation method into the validation plan, and test marked glass samples against exactly that regime.

A workable verification routine, in the spirit of ISO 28219, looks like this:

  1. Grade the code on freshly marked samples against the 1.5 © minimum, using the specified aperture and angle, rather than judging by eye.
  2. Run the samples through the full reprocessing regime for the intended number of cycles.
  3. Grade again and compare. Record both results in the design file.
  4. Check the glass: inspect for microcracks, measure strength where the marked zone is load-bearing, and confirm coatings in the area are intact.
  5. Repeat on a different production lot, so that the result reflects the process and not one lucky batch.

The scanning side is a manufacturing question too. Placement of the code, contrast and lighting on the production line decide how often a code is rejected. These panel labelling checks show the kind of barcode and scanning checks a production line benefits from, and the same discipline applies to glass.

Glass finishes also interact with legibility. An anti-reflective coating changes contrast and an oleophobic layer changes how a code reads under a handheld imager. Read our guide to cleaning AR coated glass when you set the reprocessing regime, and our guide to biocompatible glass and ISO 10993 actions when the marked surface touches the patient.

What to put on the glass drawing and RFQ

A glass supplier can only make a mark-ready part if the request says what the mark must do. Include:

  1. Marking zone. Position, size and clear area, with a note on whether the zone is on a coated, printed or bare surface.
  2. Code and size. Symbology (Data Matrix), module size, and the minimum grade the device maker will accept.
  3. Method and stage. Laser etch or printed mark, and whether it happens before or after strengthening, coating or printing.
  4. Reprocessing life. Number of cycles, cleaning chemistry, temperature and sterilisation route.
  5. Traceability. The lot information that must be recorded for each delivered part and kept for the file.
  6. Documentation. Inspection records, material certificates and any export compliance paperwork the device file needs.

With those points set, the RFQ for CNC glass machining or for optical windows can be priced and checked against one clear specification, and prototype and volume lots stay consistent. If you are unsure where the line falls between the engineering of the glass and the regulatory duty of the device, our medical device glass engineer’s guide and terminology guide give the shared vocabulary.

Our view on UDI and glass components

We see UDI as a traceability question before it is a marking question. The device maker owns the identifier, and the glass underneath has to be traceable to the same standard, lot by lot, through every cutting, machining, finishing and coating step.

That is why we work with ISO 9001 certification, traceable production and batch-level traceability as a matter of course, and why our quality control uses advanced inspection. A device file is only as strong as its weakest record.

Our advice to engineering teams is to put the mark on the drawing early. Decide the zone, the method, the order of operations and the reprocessing life together with the strengthening, coating and print steps, because those choices cannot be separated on glass. Scaling from prototype to volume is simpler when the specification already says what the mark must survive.

We would rather agree that specification at the start of a project than discover at validation that the marking zone and the coating compete for the same surface.

Alexandra

Specify mark-ready glass with us

Precision Glasses makes custom glass components for medical devices, defence, aerospace, automotive dashboards, lighting, electronics and marine technology. Our services include glass cutting, CNC glass machining, edge finishing, toughened and laminated glass, AR and AG coated glass, display glass, optical windows and covers, and custom components, with ceramic printing and documented export compliance. Production is traceable and certified to ISO 9001.

Precision Glasses

Send us the drawing, the volumes and the reprocessing requirements, and we will work through the marking zone and the order of operations with your team. Start at glassprecision.com.

FAQ

Does a glass component need its own UDI?

Normally not. GOV.UK says raw materials, component parts and intermediate products are not normally medical devices, because the medical purpose relates to the finished product. The finished device carries the UDI, and the glass supplier provides lot traceability that supports it.

Is UDI marking mandatory in Great Britain?

Not yet under the Medical Devices Regulations 2002, where UDI information is requested by the MHRA on a voluntary basis at registration. A draft amendment summarised in May 2026 would make it mandatory, so check the MHRA and GOV.UK for the final rules and transition dates. Northern Ireland follows EU MDR, which already contains UDI rules.

Why is dot peen not used on glass?

Dot peen uses a stylus that needs a hard, deformable material. GS1 Healthcare Japan’s guide says it is unsuitable for glass and recommends laser etching for small instruments. Whichever method you choose, validate it against the device’s reprocessing regime.

Which barcode should be marked directly on a glass part?

A matrix code such as Data Matrix. ISO 28219 limits direct part marking to matrix symbologies and sets a minimum symbol grade of 1.5 ©. Smaller modules need special imagers, so choose the code size and the reader together.

How long must a direct mark stay readable?

For the intended lifetime of the device, including after each cleaning, disinfection or sterilisation process, according to EU MDR Annex VI Part C. The IMDRF guide reports that a properly validated laser-marked Data Matrix stayed readable after at least 500 reprocessing cycles, and that alkaline cleaners fade marks faster.